Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikačního výsledku
DREXLER, P.; NEŠPOR, D.; FIALA, P.; BARTUŠEK, K.; KUBÁSEK, R.
Originální název
Design and Fabrication of Planar Magnetoinductive Resonator Arrays for MRI System Field Shaping
Anglický název
Druh
Článek recenzovaný mimo WoS a Scopus
Originální abstrakt
For the experiments with MR system sensitivity increasing by means of periodical metamaterial structures a magnetoinductive resonator array has to be designed. Proposed design guidelines for precise new array assembly are described and the results of new arrays measurement are presented. For the array's resonant frequency measurement the dip meter method is described. For resonant frequency measurement of single loop a return loss method is presented. As a suitable alternative a new coaxial shunt measurement method is proposed. The results of the new arrays measurement by means of coaxial shunt method are presented also.
Anglický abstrakt
Klíčová slova
magnetoinductive resonator array, magnetic resonance
Klíčová slova v angličtině
Autoři
Rok RIV
2012
Vydáno
20.03.2011
Nakladatel
PIERS
ISSN
1559-9450
Periodikum
Progress In Electromagnetics
Svazek
2011
Číslo
Stát
Spojené státy americké
Strany od
643
Strany do
647
Strany počet
5
BibTex
@article{BUT30839, author="Petr {Drexler} and Dušan {Nešpor} and Pavel {Fiala} and Karel {Bartušek} and Radek {Kubásek}", title="Design and Fabrication of Planar Magnetoinductive Resonator Arrays for MRI System Field Shaping", journal="Progress In Electromagnetics", year="2011", volume="2011", number="2011", pages="643--647", issn="1559-9450" }