Originální abstrakt
The paper present new approach in the field of precise electrolytic conductivity measurements with planar sensors. This novel measuring method was suggested for measurement with IDEs thick-film electrodes. Correction characteristics over a wide range of a specific conductivities were determined from an interface impedance characterisation of the thick-film IDEs. A local minimum of the imaginary part of the interface impedance is used for corrections to get linear responses. The ridge frequency was determined at wide range of measured conductivity. An iteration mode of measurements was suggested to exactly measure the conductivity in the ridge frequency in order to achieve a highly accurate response. The method takes precise conductivity measurements in concentration ranges from 10-6 to 1 M without cell replacement.
Anglický abstrakt
Klíčová slova
IDEs, electrolytic conductivity, iterative measurement, ridge frequency
Klíčová slova v angličtině
Autoři
Vydáno
01.12.2002
Nakladatel
Czech Technical UniversityFaculty of Electrical EngineeringDepartment of MeasurementTechnická 2, Prague 6-Dejvice166 27 Czech Republic
Místo
FEL CVUT Praha
ISBN
80-01-02576-4
Kniha
Eurosensors XVI, The 16th European Conference on Solid-State Transducers
Strany od
602
Strany počet
2
URL
UMEL
BibTex
@inproceedings{BUT4955, author="Jaromír {Hubálek} and Jan {Krejčí}", title="Wide Range Precise Conductivity Measurement without Cell Replacement", booktitle="Eurosensors XVI, The 16th European Conference on Solid-State Transducers", year="2002", pages="2", publisher="Czech Technical UniversityFaculty of Electrical EngineeringDepartment of MeasurementTechnická 2, Prague 6-Dejvice166 27 Czech Republic", address="FEL CVUT Praha", isbn="80-01-02576-4", url="UMEL" }