Cross-wire Calibration for Freehand 3D Ultrasonography: Measurement and Numerical Issues

Asterios ANAGNOSTOUDIS, Jiří JAN

Originální název

Cross-wire Calibration for Freehand 3D Ultrasonography: Measurement and Numerical Issues

Anglický název

Cross-wire Calibration for Freehand 3D Ultrasonography: Measurement and Numerical Issues

Druh

Článek recenzovaný mimo WoS a Scopus

Originální abstrakt

3D freehand ultrasound is an imaging technique, which is gradually finding clinical applications. A position sensor is attached to a conventional ultrasound probe, so that B-scans are acquired along with their relative locations. This allows the B-scans to be inserted into a 3D regular voxel array, which can then be visualized using arbitrary-plane slicing, and volume or surface rendering. A key requirement for correct reconstruction is the calibration: determining the position and orientation of the B-scans with respect to the position sensor’s receiver. Following calibration, interpolation in the set of irregularly spaced B-scans is required to reconstruct a regular-voxel array. This text describes a freehand measurement of 2D ultrasonic data, an approach to the calibration problem and several numerical issues concerned with the calibration and reconstruction.

Anglický abstrakt

3D freehand ultrasound is an imaging technique, which is gradually finding clinical applications. A position sensor is attached to a conventional ultrasound probe, so that B-scans are acquired along with their relative locations. This allows the B-scans to be inserted into a 3D regular voxel array, which can then be visualized using arbitrary-plane slicing, and volume or surface rendering. A key requirement for correct reconstruction is the calibration: determining the position and orientation of the B-scans with respect to the position sensor’s receiver. Following calibration, interpolation in the set of irregularly spaced B-scans is required to reconstruct a regular-voxel array. This text describes a freehand measurement of 2D ultrasonic data, an approach to the calibration problem and several numerical issues concerned with the calibration and reconstruction.

Klíčová slova v angličtině

3D freehand ultrasonography, cross-wire calibration phantom, 3D freehand ultrasound calibration

Autoři

Asterios ANAGNOSTOUDIS, Jiří JAN

Vydáno

01.07.2005

Místo

Brno

Edice

Radioengineering

ISSN

1210-2512

Periodikum

Radioengineering

Svazek

2005

Číslo

2

Stát

Česká republika

Strany od

222

Strany počet

6

BibTex

@article{BUT46273,
  author="Asterios {Anagnostoudis} and Jiří {Jan}",
  title="Cross-wire Calibration for Freehand 3D Ultrasonography: Measurement and Numerical Issues",
  journal="Radioengineering",
  year="2005",
  volume="2005",
  number="2",
  pages="6",
  issn="1210-2512"
}