Detail publikačního výsledku

Implantable Medical Device for Measuring Vital Signs of Animals - Proof of Concept

LENGÁL, O.; HÁZE, J.

Originální název

Implantable Medical Device for Measuring Vital Signs of Animals - Proof of Concept

Anglický název

Implantable Medical Device for Measuring Vital Signs of Animals - Proof of Concept

Druh

Stať ve sborníku v databázi WoS či Scopus

Originální abstrakt

This paper presents the design of medical device implantable under the skin of animals for vital signs measurement (temperature and heartbeat) and wireless data distribution for health data review. This system utilizes a thermoelectric generator as an energy harvester and supercapacitor as an energy storage for long term operation without the need of system maintenance and Bluetooth Low Energy module for data transfer to any Bluetooth compatible device. Design also includes passive RFID tag for identification. The system is presented as proof of concept with commercially available electronic components.

Anglický abstrakt

This paper presents the design of medical device implantable under the skin of animals for vital signs measurement (temperature and heartbeat) and wireless data distribution for health data review. This system utilizes a thermoelectric generator as an energy harvester and supercapacitor as an energy storage for long term operation without the need of system maintenance and Bluetooth Low Energy module for data transfer to any Bluetooth compatible device. Design also includes passive RFID tag for identification. The system is presented as proof of concept with commercially available electronic components.

Klíčová slova

battery management; Bluetooth Low Energy; DC-DC converter; energy harvesting; proof of concept; supercapacitor; vital sign measurement

Klíčová slova v angličtině

battery management; Bluetooth Low Energy; DC-DC converter; energy harvesting; proof of concept; supercapacitor; vital sign measurement

Autoři

LENGÁL, O.; HÁZE, J.

Rok RIV

2026

Vydáno

29.04.2025

Nakladatel

Brno University of Technology

Místo

Brno

ISBN

9788021463219

Kniha

Proceedings II of the 31st Student EEICT 2025

Strany od

138

Strany do

141

Strany počet

4

URL

BibTex

@inproceedings{BUT201516,
  author="Ondřej {Lengál} and Jiří {Háze}",
  title="Implantable Medical Device for Measuring Vital Signs of Animals - Proof of Concept",
  booktitle="Proceedings II of the 31st Student EEICT 2025",
  year="2025",
  pages="4",
  publisher="Brno University of Technology",
  address="Brno",
  isbn="9788021463219",
  url="https://www.scopus.com/pages/publications/105012494182?origin=resultslist#"
}