BARDOŇOVÁ, J.; PROVAZNÍK, I.; NOVÁKOVÁ, M.; NOGOVÁ, K.; SEKORA, J.
Original Title
New Recording Setup for Ratiometric Recording of Action Potentials by Optical Means
English Title
Type
Paper in proceedings (conference paper)
Original Abstract
Monophasic action potential (MAP) can be recorded from the heart surface by optical method via fluorescence measurement. The motion of isolated heart during experiment caused additional noise in recorded signal. The motion artifact can be eliminated by ratiometric fluorescence emission measurements. This study is based on experiments in which a new ratiometric recording setup is used for recording MAPs from isolated rabbit hearts perfused according to Langendorff. MAPs recorded by new ratiometric recording setup are compared with simple optical MAP measurement without ratiometry.
English abstract
Keywords
Recording, Ratiometric Recording, Action Potentials
Key words in English
Authors
RIV year
2010
Released
01.12.2008
Publisher
IEEE
Location
Bologna, Italy
ISBN
0-7803-8894-1
Book
Computers in Cardiology
Pages from
1
Pages to
4
Pages count
Full text in the Digital Library
http://hdl.handle.net/
BibTex
@inproceedings{BUT27046, author="Jana {Kolářová} and Valentýna {Provazník} and Marie {Nováková} and Kateřina {Nogová} and Jiří {Sekora}", title="New Recording Setup for Ratiometric Recording of Action Potentials by Optical Means", booktitle="Computers in Cardiology", year="2008", pages="1--4", publisher="IEEE", address="Bologna, Italy", isbn="0-7803-8894-1" }