GESCHEIDTOVÁ, E.; BARTUŠEK, K.
Original Title
Measurement of Radio Frequency Magnetic Field
English Title
Type
Peer-reviewed article not indexed in WoS or Scopus
Original Abstract
A high quality of magnetic fields is one of the basic prerequisites for a perfect operation of the MR tomographic scanner. Magnetic susceptibility of materials that are used for body implants causes much distortion in MR images. The paper deals with mapping the radiofrequency magnetic field induction in the vicinity of these materials with the aim of characterizing the magnetic field and comparing the results with fields simulated in the ANSYS program. Magnetic resonance method of radiofrequency magnetic field mapping was verified and optimized for both the characterizing of MR imaging detecting coil and the effects of electrical conducting samples. The designed method is based on optimized spin-echo imaging sequence with small flip angle.
English abstract
Keywords
MR tomographic scanner, magnetic susceptibility, body implants, mapping the radiofrequency magnetic field, ANSYS program.
Key words in English
Authors
Released
20.03.2007
Publisher
The electromagnetic academy
Location
Cambridge
ISBN
1931-7360
Periodical
PIERS ONLINE
Volume
2
Number
6
State
United States of America
Pages from
555
Pages to
558
Pages count
4
BibTex
@article{BUT45363, author="Eva {Gescheidtová} and Karel {Bartušek}", title="Measurement of Radio Frequency Magnetic Field", journal="PIERS ONLINE", year="2007", volume="2", number="6", pages="555--558", issn="1931-7360" }