HOVANCOVÁ, J.; IVANA, Š.; VANÝSEK, P.; ORINAKOVÁ, R.; SHEPA, I.; KAŇUCHOVÁ, M.; KIRÁLY, N.; VOJTKO, M.; ČUDEK, P.; ORIŇAK, A.
Original Title
Ligand-to-metal charge transfer (LMCT) complex: New approach to non- enzymatic glucose sensors based on TiO2
English Title
Type
WoS Article
Original Abstract
Titanium metal covered with titanium dioxide (Ti/TiO2) was used as a promising non-enzymatic glucose sensor demonstrating a new approach to electrochemical glucose determination. A Ti/TiO2 nanotubes-glucose complex formation on an electrode surface was demonstrated by infrared, Raman, and X-ray photoelectron spectroscopy. The main electrochemical method used for glucose detection was electrochemical impedance spectroscopy as a promising method for an easy electrochemical determination. To the best of our knowledge, this kind of sensor has never been studied especially in case of non-enzymatic glucose detection. Ti/TiO2 glucose sensor displays a wide practical linear range (1–15 mM) in a pH neutral condition, which cover the typical range in diabetic patients. The selectivity and the limit of detection were 8.3 kΩ·mM−1 and 1.2 mM, respectively. Therefore this kind of a system is very promising candidate for blood glucose sensing which should be able to replace enzymatic glucose sensors. Based on the properties of the prepared sensors, TiO2 is an unrivalled material for glucose sensing comparable with the enzymatic sensors.
English abstract
Keywords
Non-enzymatic glucose sensor; Ligand-to-metal charge transfer complex; Titanium dioxide glucose sensors
Key words in English
Authors
RIV year
2021
Released
21.08.2020
Publisher
Elsevier
ISBN
1572-6657
Periodical
Journal of Electroanalytical Chemistry
Volume
878
Number
1 December
State
Swiss Confederation
Pages from
1
Pages to
8
Pages count
URL
https://www.sciencedirect.com/science/article/pii/S1572665720308171
BibTex
@article{BUT164854, author="Jana {Hovancová} and Šišoláková {Ivana} and Petr {Vanýsek} and Renáta {Orinaková} and Ivan {Shepa} and Mária {Kaňuchová} and Nikolas {Király} and Marek {Vojtko} and Pavel {Čudek} and Andrej {Oriňak}", title="Ligand-to-metal charge transfer (LMCT) complex: New approach to non- enzymatic glucose sensors based on TiO2", journal="Journal of Electroanalytical Chemistry", year="2020", volume="878", number="1 December", pages="1--8", doi="10.1016/j.jelechem.2020.114589", issn="1572-6657", url="https://www.sciencedirect.com/science/article/pii/S1572665720308171" }
Documents
1-s2.0-S1572665720308171-main