Publication detail

Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies

FUIS, V.

Original Title

Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies

English Title

Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies

Type

conference paper

Language

en

Original Abstract

The paper presents the results of the computational modelling of the stress and reliability of the ceramic head of the total hip joint endoprosthesis under ISO 7206-5 loading. Endoprosthesis consists of the two main parts – ceramic head and the steel stem. The joint of these two parts is realised by the self-locking cone and the manufacturing inaccuracies of these cones significantly influence the tensile stresses in the head and head’s failure probability. The manufacturing inaccuracies were measured on 3D measuring device, stress in the analysed system is calculated using FEM system ANSYS and the head’s reliability calculation is based in the Weibull weakest link theory.

English abstract

The paper presents the results of the computational modelling of the stress and reliability of the ceramic head of the total hip joint endoprosthesis under ISO 7206-5 loading. Endoprosthesis consists of the two main parts – ceramic head and the steel stem. The joint of these two parts is realised by the self-locking cone and the manufacturing inaccuracies of these cones significantly influence the tensile stresses in the head and head’s failure probability. The manufacturing inaccuracies were measured on 3D measuring device, stress in the analysed system is calculated using FEM system ANSYS and the head’s reliability calculation is based in the Weibull weakest link theory.

RIV year

2003

Released

24.06.2003

Publisher

Politechnika Slaska, Katedra Mechaniky Stosowanej, Gliwice

Location

Jaworzynka, Poland

ISBN

83-917224-3-0

Book

Applied Mechanics 2003

Pages from

51

Pages to

56

Pages count

6

Documents

BibTex


@inproceedings{BUT8179,
  author="Vladimír {Fuis}",
  title="Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies",
  annote="The paper presents the results of the computational modelling of the stress and reliability of the ceramic head of the total hip joint endoprosthesis under ISO 7206-5 loading. Endoprosthesis consists of the two main parts – ceramic head and the steel stem. The joint of these two parts is realised by the self-locking cone and the manufacturing inaccuracies of these cones significantly influence the tensile stresses in the head and head’s failure probability. The manufacturing inaccuracies were measured on 3D measuring device, stress in the analysed system is calculated using FEM system ANSYS and the head’s reliability calculation is based in the Weibull weakest link theory.",
  address="Politechnika Slaska, Katedra Mechaniky Stosowanej, Gliwice",
  booktitle="Applied Mechanics 2003",
  chapter="8179",
  institution="Politechnika Slaska, Katedra Mechaniky Stosowanej, Gliwice",
  year="2003",
  month="june",
  pages="51",
  publisher="Politechnika Slaska, Katedra Mechaniky Stosowanej, Gliwice",
  type="conference paper"
}