Přístupnostní navigace
E-application
Search Search Close
Publication result detail
KUČÍREK, P.; ŠIKULA, O.
Original Title
Selection of a suitable C02 sensor for indoor air quality assessment
English Title
Type
Paper in proceedings outside WoS and Scopus
Original Abstract
This article compares eight types of CO2 sensors using NDIR technology for accuracy, response time, and acquisition costs. The experiment included static and dynamic tests in a chamber and a duct. The Testo 935 and Rotronic CL11/CP11 sensors showed high accuracy and fast response, while the Winsen MH-Z16 offered the best performance-to-cost ratio. Other tested sensors – NetAtmo, Comet S3532, LaskaKit SCD41 and MH-Z14 exhibited inaccuracies, making them unsuitable for scientific applications.
English abstract
Keywords
CO2 measurement, CO2 sensors, pollutant measurement, experimental sensor comparison, accuracy, response time
Key words in English
Authors
RIV year
2026
Released
05.12.2025
Publisher
Brno University of Technology, Faculty of Civil Engineering
Location
Brno
Pages from
1
Pages to
11
Pages count
BibTex
@inproceedings{BUT199899, author="{} and Pavel {Kučírek} and Ondřej {Šikula}", title="Selection of a suitable C02 sensor for indoor air quality assessment", year="2025", pages="1--11", publisher="Brno University of Technology, Faculty of Civil Engineering", address="Brno", doi="10.13164/juniorstav.2025.26095" }